Validation vs verification in the laboratory Two words used interchangeably for two jobs of very different size. Getting the distinction right is the difference between a week of work and a month, or between passing an inspection and failing one.

Before a laboratory can report a patient result from a new method, it has to show the method performs acceptably on its own instruments, with its own staff. How much work that takes comes down to one question, and the question is easy to get wrong. Are you verifying a method someone else has already characterised? Or validating one whose performance you have to establish yourself?

Answer wrong in one direction and you waste weeks proving what the manufacturer already documented. Answer wrong in the other and you put a method into service that you never properly evaluated.

The distinction

Verification is confirming that a method whose performance has already been established lives up to its stated claims on your instruments, with your staff, in your environment. The typical case is an FDA-cleared or CE-marked assay, used exactly as the manufacturer intended. The manufacturer did the heavy characterisation during development. You are checking that it holds.

Validation is establishing performance characteristics from scratch, because no one has done it for your intended use. Validation applies to a laboratory-developed test, an assay you have modified or one you are using outside its cleared indications. Here there are no claims to confirm. You have to generate the evidence yourself, across a wider set of characteristics and larger studies.

The distinction is simple: verification confirms someone else’s numbers, and validation produces your own. Verification is a fraction of the work, but only if you are entitled to it.

Which one you need

Your method is… You must…
FDA-cleared or CE-marked, used unmodified per the instructions for use Verify the manufacturer’s claims
Laboratory-developed, modified from its cleared use, or used off-label Validate — establish performance from scratch

The deciding factor is whether the assay is used exactly as cleared. A cleared assay run on a different specimen type, at a different dilution or on an unapproved instrument has been modified. Modification moves you from verification to validation. When in doubt, treat any change from the cleared use as a modification.

A decision fork. The question is whether the assay is FDA-cleared or CE-marked and used exactly as the manufacturer intended. Yes leads to VERIFY, with four characteristics: precision, trueness, reportable range and a transferred reference interval, taking days. No leads to VALIDATE, with seven characteristics adding analytical sensitivity, analytical specificity and an established reference interval, taking weeks.
One question decides the size of the job. The verify branch confirms four characteristics the manufacturer has already established. The validate branch establishes seven from scratch. A different specimen type, dilution or instrument moves you from the first to the second, so when in doubt, validate.

What verification involves

A verification confirms the claims that matter for reporting results: precision, trueness, the reportable range and the reference interval. The work is deliberately compact, a handful of days rather than weeks, because you are checking established numbers rather than deriving them. In practice that means:

  • Verifying precision and trueness against the manufacturer’s claims (EP15-A3 from the Clinical and Laboratory Standards Institute, CLSI): a short repeatability and within-laboratory study tested against the claimed imprecision. The same protocol estimates bias against materials with assigned values.
  • Confirming the agreement with the existing method where you are replacing one (CLSI EP09-A3), so patient results stay comparable across the changeover.
  • Verifying the reportable range with materials at the low, middle and high end of the claimed range (CLSI EP06-Ed2 where a full linearity check is needed).
  • Verifying that the reference interval transfers to your population (CLSI EP28-A3c): a small study of around 20 reference individuals rather than a full establishment.

What validation adds

Validation covers everything verification does, but establishes each characteristic rather than confirming it. Validation also adds the analytical characteristics a manufacturer would have generated during development:

  • Establishing precision across a full nested design of days and runs (CLSI EP05-A3), rather than the short verification study.
  • Establishing the detection capability (limit of blank, limit of detection and limit of quantitation; CLSI EP17-A2) where low-end performance matters.
  • Establishing the measuring interval over which the response is linear (CLSI EP06-Ed2).
  • Establishing analytical specificity: interference from endogenous substances and drugs (CLSI EP07-Ed3).
  • Establishing the reference interval from scratch (CLSI EP28-A3c), typically 120 or more reference individuals per partition.
  • For a diagnostic claim, establishing diagnostic performance (receiver operating characteristic curves; CLSI EP24-A2) or sensitivity and specificity (CLSI EP12-A2) against an appropriate reference.

The regulatory backdrop

The distinction is not just good practice. The frameworks laboratories are inspected against are written around it.

Under the Clinical Laboratory Improvement Amendments (CLIA) in the US, an unmodified FDA-cleared assay requires verification of accuracy, precision, reportable range and reference intervals before reporting. A modified or laboratory-developed high-complexity test requires more: establishing analytical sensitivity, analytical specificity and other characteristics as well. ISO 15189 and the checklists of the College of American Pathologists (CAP) draw the same line.

Which side of that line your method falls on determines what an inspector will expect to see in your file. Settle the question before you start collecting data, not after.

Common mistakes

Validating an unmodified cleared assay. Running a full establishment study for a method the manufacturer has already characterised is weeks of effort spent confirming documented claims. Verify unless something about your use is non-standard.

Verifying when you should have validated. The more dangerous error: treating a modified or laboratory-developed test as if a short verification suffices. If you changed the specimen, the platform or the intended use, the manufacturer’s claims no longer apply and you need a full validation.

Skipping the reference interval. Precision and trueness get the attention, but reporting a manufacturer’s reference interval without at least verifying it transfers to your population is a common inspection finding.

Deciding after the fact. Whether you are validating or verifying determines your study design and sample sizes. Settle it first. Discovering mid-study that you under-powered a validation is expensive.

Do either with Analyse-it

Analyse-it performs the statistics for most of the validation and verification studies against the CLSI EP protocols, inside Excel:

Every feature from all five editions for 15 days. These are in the Method Validation and Ultimate editions, from US$ 475 a year. Go to the guide for the task in front of you once you know which path you are on. The task may be verifying precision claims, choosing a method-comparison regression, reference intervals, detection capability or diagnostic performance. For the regulatory framing, see what CLIA requires before you report a result or method validation under ICH Q2(R2). Either path ends in a document: what goes in a validation report.