Statistical software for clinical laboratories Verify a manufacturer’s claims before you report a patient result and keep the method in control afterwards, with the evidence in a workbook you can hand to an inspector.

Every feature from all five editions for 15 days, no sign-up.
Method Validation edition from US$ 475 a year · 30-day money-back guarantee.

Microsoft Excel with the Analyse-it tab of the Method Validation edition selected: a method comparison report with the scatter plot, the ordinary least squares fit with its confidence interval and the allowable difference bands, the Method Comparison task pane open on the fit options, and the Method Comparison menu dropped open on the ribbon listing the plots, bias estimators and agreement estimators. Handwritten notes: Runs inside Excel, and every edition includes the statistics research needs:; Method Validation adds method comparison, MSA - trueness, precision, linearity - diagnostic accuracy and reference intervals; Linear fit with its 95% CI and the allowable difference: plots and tables, on a worksheet; Every option for the analysis is set here, then Recalculate; The output is an Excel worksheet: share it with colleagues, auditors or regulators, archive it, open it on any PC with Excel.
I started working with Analyse-it about 15 years ago. I was then working for a medical laboratory in method evaluation and ISO certification. The certification expert was delighted that I use the same statistical tool that he uses. Later in the pharma industry I applied it again to method evaluation and statistical work to analyse our projects. Now I have a company for clinical trials. Even tasks with ten thousands of numbers are no problem with Analyse-it.
Peter J.
Clinical Immunology FAMH

Most of what a clinical laboratory does statistically is verification. The manufacturer has already established the performance claims. Your job is to demonstrate that you reproduce them in your hands, on your analysers, with your staff and your patient population. A verification is a much smaller study than the one the manufacturer ran. The study still has to be done before you report a result, repeated when something changes and evidenced when an inspector asks.

The awkward part is rarely the statistics. The difficulty is that the requirement lives in one document, the acceptance limit in another and the calculation in a spreadsheet somebody built years ago. Analyse-it runs every one of these studies directly in Excel, so the analysis, the data and the record are one workbook. The workbook is one you can hand to an assessor, archive and reopen in three years when the same question comes back. Follow a CLSI protocol exactly where that is what your accreditation body expects, or design the study to suit your method and use the statistics directly. The software does not decide that for you.

Trusted for nearly 30 years by 75,000 scientists in clinical, research and quality control laboratories.

Mayo Clinic Roche Memorial Sloan Kettering Cancer Center Siemens Healthineers CHUV Lausanne University Hospital Sysmex Odense University Hospital North Dakota Department of Health

What CLIA, CAP and ISO 15189 require, and which study shows it

The requirements below are the ones that come up in almost every accredited laboratory. Each links to a guide covering how the study is designed, how to set the acceptance limit and what usually goes wrong. The guides are useful whichever software you end up using.

What you must show The study Typically required by Guide
Precision Short replication study tested against the claimed imprecision (EP15-A3) CLIA, CAP, ISO 15189 Verifying a precision claim
Trueness or accuracy Comparison against a reference or comparative method, with bias at the decision points CLIA, CAP, ISO 15189 Bias at a decision point
Reportable range — the AMR Linearity across the claimed interval, repeated every six months under CAP CLIA, CAP AMR verification
Reference intervals Transfer the manufacturer’s interval and verify it with a binomial test on a small sample (EP28-A3c) CLIA, CAP, ISO 15189 Transferring a reference interval
Analytical sensitivity — LoB, LoD, LoQ Only if you modified the method or it is laboratory-developed (EP17-A2) CLIA, for modified and LDT methods LoB, LoD and LoQ explained
Analysers agreeing with each other Difference study between instruments running the same method, twice a year CAP Comparing instruments
The method staying in control Daily internal QC on Levey-Jennings charts with a rule set CLIA, ISO 15189 Levey-Jennings charts
Measurement uncertainty Built top-down from the imprecision and bias your verification already produced ISO 15189 Top-down uncertainty

Not sure which of these apply to your laboratory? What CLIA requires before you report a result works through the two lists in 42 CFR 493.1253 and which one applies to your method.

Validation or verification: which one you are doing

An unmodified FDA-cleared assay run exactly as the manufacturer intends needs four characteristics verified. Modify it — a different sample type, a different matrix, an off-label specimen — and it becomes laboratory-developed. The second CLIA list then applies, and you are establishing performance rather than confirming it. Laboratories routinely discover this after the study is finished and the paperwork is written.

The distinction is not about how careful you are, and it is not negotiable at inspection. Validation versus verification sets out how to tell which one you are on, and verifying a manufacturer’s precision claim covers the difference in study design that follows from it.

Verify precision and trueness before reporting

The EP15-A3 verification is deliberately small — typically five days, five replicates, two or three concentrations — which is exactly why the statistics matter. With that few observations, a point estimate of imprecision that merely looks smaller than the claim proves very little. The protocol tests the observed imprecision against the claim rather than judging it by eye, and lets you estimate bias from the same runs rather than mounting a second study.

Analyse-it runs the verification as one analysis: within-run and total precision with χ² tests against the manufacturer’s claimed imprecision, and trueness against assigned values at the same time. When a verification fails, the output separates the run-to-run component from the replicate scatter. That separation is usually what tells you whether the problem is the method, the operator or the material.

Precision and trueness in detail →
The EP15-A3 ferritin precision verification example report: page one of the PDF as Analyse-it produces it. 3 pages EP15-A3 — Table 8
Ferritin precision verification.
3 samples × 5 runs × 5 observations. Outlier identification with the flagged results excluded, then a variability plot and within-run and total precision. A χ² test of each against the manufacturer’s claim at a 1.67% individual significance level.
The EP15-A3 ferritin trueness verification example report: page one of the PDF as Analyse-it produces it. 2 pages EP15-A3 — Example 1Z
Ferritin trueness verification.
5 runs × 5 observations against a proficiency testing sample with an assigned value of 142.5 and a stated uncertainty. Recovery, bias with a 98.3% CI against an allowable bias and a t test of the bias.

Confirm the reportable range and the reference intervals

AMR verification and calibration verification answer different questions, yet they are conflated more often than any other pair of studies in the CAP checklists. Calibration verification asks whether the analyser still reads correctly at the calibration points. AMR verification asks whether the response stays linear across the whole claimed interval. Only the AMR verification tells you what you may report without dilution.

For reference intervals, transferring the manufacturer’s is usually defensible and far cheaper than recruiting a reference population. Transference is still a study with a result, not an assumption. Analyse-it verifies a transferred interval with the binomial test on a small sample per EP28-A3c. Intervals can also be estimated outright, partitioned by sex or age, where the transfer does not hold.

Linearity and measuring interval →   Reference intervals →
The EP28-A3c reference interval transference example report: page one of the PDF as Analyse-it produces it. 2 pages EP28-A3c
Mercury reference interval transference.
20 observations per sex, tested against limits transferred from another laboratory. The proportion of results falling inside the transferred interval, with an exact binomial test against 95%.
The calcium linearity example report: nonlinearity at each dilution with its confidence interval against the allowable band, on page two of the PDF as Analyse-it produces it. 6 pages EP06-A — Appendix C
Calcium linearity, full and reduced interval.
6 dilutions × 2 observations. Over the full interval five of the six dilutions exceed the ±0.2 mg/dL allowable nonlinearity, so a second analysis refits over a reduced interval that passes throughout. Linear and polynomial fits with nonlinearity at each dilution.

Method comparison: show your analysers agree with each other

Two analysers running the same method on the same bench is not a method comparison. Treating it as one is the most common analytical error in accreditation work. No reference method exists here, neither instrument is “right”, and fitting a regression to decide which is biased answers a question nobody asked. What matters is the size and spread of the differences, judged against a limit you set before you looked at the data.

Analyse-it produces the difference plot, mean and median bias with confidence intervals, and limits of agreement, each judged against the allowable difference you set. This output is the evidence the CAP twice-yearly comparison asks for.

Agreement and difference plots →
The EP21-A LDL cholesterol total analytical error example report: page one of the PDF as Analyse-it produces it. 2 pages EP21-A — Table 2
LDL cholesterol total analytical error.
100 observations. Difference plot and mountain plot with an allowable difference of ±10 mg/dL, median difference with a 90% CI and 95% limits of agreement.
The EP09-A3 partitioned difference plots example report: page one of the PDF as Analyse-it produces it. 2 pages EP09-A3 — Appendix I
Difference plots over a partitioned interval.
79 observations split at 1.8 µg/L, 40 below and 39 above. Each interval gets its own difference plot, mean difference with a 95% CI and a hypothesis test. The allowable difference is ±0.06 µg/L below the split and ±6% above it. Neither interval shows a significant bias.

Keep the method in control between verifications

A verification is a snapshot. What protects patient results for the rest of the year is the internal QC chart. What makes the chart work or fail is the rule set. Too few rules and a slow reagent drift runs for weeks before a point crosses a limit. Too many and staff learn to dismiss the flags, which is worse than having none.

Levey-Jennings charts are Shewhart individuals charts with clinical labelling. Analyse-it draws them with the same detection rule sets, phases before and after a lot change, and stratification by analyser or shift. Where a drift is small and sustained rather than sudden, CUSUM and EWMA catch it earlier than any Shewhart rule will. Reading out-of-control signals covers how to tell a real shift from the false alarms any rule set produces.

Control charts and detection rules →
The Shewhart, EWMA and CUSUM example report: the EWMA chart with its tapering control limits, on page three of the PDF as Analyse-it produces it. 6 pages Shewhart, EWMA and CUSUM
Copper concentration in a plating pool.
Three analyses on one dataset across the IQ, OQ and production phases. Xbar-R with Montgomery rules 1, 2, 3, 5 and 8 and points coloured by operator, flagging seven signals. EWMA with λ = 0.2 and L = 3 flags eight, and CUSUM with h = 5 and k = 0.5 flags eleven. Every signal falls in the IQ phase but one.
The individual and moving range example report: page one of the PDF as Analyse-it produces it. 2 pages Individual and moving range
Loan processing costs.
Montgomery, page 268. 40 observations over two phases with a known mean and sigma. I-MR chart with ±3 sigma limits, flagging observation 39 on both charts and observation 40 on the individuals chart.

Where the acceptance limits come from

None of the regulations set your numbers. CLIA names the characteristics you must verify; CAP tells you how often; ISO 15189 requires that you have criteria and apply them. Choosing the limit is the laboratory’s job. That choice is also the one most often made after the data has been collected — which is precisely when it stops being defensible.

The Milan hierarchy is the usual starting point. The hierarchy runs from clinical outcome where the evidence exists, through biological variation where it does not, to the state of the art as the fallback. From there, allowable total error and total error versus measurement uncertainty cover the two competing ways of combining imprecision and bias into one number you can test against.

For ISO 15189 laboratories, the uncertainty estimate does not need a new study: the top-down approach builds it from the imprecision and bias your verification has already produced.

Which edition a laboratory needs

The verification studies above are the Method Validation edition. That edition holds precision and trueness against a claim, method and instrument comparison, linearity and the AMR, detection limits and reference intervals. Daily internal QC on Levey-Jennings charts is the Quality Control & Improvement edition, and Ultimate holds both in one licence.

Every edition includes the full Standard edition statistics, and the comparison table lists every analysis against every edition. Volume pricing is available for laboratory networks.

Every calculation is performed by Analyse-it — no Excel formulas and no third-party functions. The calculations are validated against the NIST Statistical Reference Datasets, CLSI reference datasets and thousands of internal test cases. That validation is the basis of the evidence you show at a CAP inspection or an ISO 15189 audit. Analyse-it runs entirely within Excel on your PC, so patient data never leaves the laboratory, and results are ordinary workbooks an assessor can open without a licence. The workbooks carry no formulas, so what you reported is what you find when you reopen the file. How Analyse-it is developed and validated →

Free trial and pricing

Try it on your own data first. The 15-day trial is every feature from all five editions, with no sign-up and no licence key — install it and start straight away.

Method Validation edition: US$ 475 per year or US$ 1155 for a perpetual licence. Every purchase carries a 30-day money-back guarantee. Need a quote for purchasing? Add the licence to the cart and save it as a PDF quote.